Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 441
Appendix H: Technical support and user maintenance
f To clean the air filter:
1. With a flat-head screwdriver turn the panel latch screw counter-clockwise
until you loosen the panel from the frame.
2. Firmly but carefully lift the panel until the white tongues are out of the frame
slots.
3. Carefully pull the panel straight back.
4. Twist off the filter housing lid thumbscrews and remove the filter housing lid.
Panel latch
Filter Filter housing lid
Filter housing lid
thumbscrew
Filter housing
5. Slide the filter from the filter housing.
6. Wash the filter with water or vacuum it to remove dust.