Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 305
Chapter 43: Acquiring Color Doppler Mode images
Acquiring Color Doppler Mode images
This chapter shows you how to acquire Color Doppler Mode images.
WARNING: High levels of ultrasound energy can damage tissue. Do not
touch the transducer when acoustic power could be generated.
In this chapter
Typical Color Doppler Mode image acquisition session ...............................................305
Color Doppler Mode window workspace .......................................................................307
Control panel controls for Color Doppler Mode.............................................................309
Color Doppler Mode acquisition settings ........................................................................313
Typical Color Doppler Mode image acquisition session
Before you begin
If you want to add physiological data to your image:
Set up your system for physiological data acquisition (page 109).
Prepare your animal on the animal platform. For detailed information refer to
the operator manual for your Vevo Imaging Station.
For blood pressure setup, see Blood Pressure section (page 113).
f To acquire a Color Doppler Mode image:
1. Press
Color
. In the image area:
The system begins storing cine loop data in the acquisition buffer
The system displays the region-of-interest (ROI) box overlay on the B-
Mode background image
If your transducer is positioned almost parallel over a vessel, the system
displays color data in the ROI box
2. To change the size and proportion of the color ROI box:
a. Press
Update
. The color ROI box becomes a dashed-line box.
Chapter 43