VDO MG 2100 Specifiche Pagina 115

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 114
Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 115
Chapter 18: Setting up to acquire physiological data
The left panel displays the physiological display setting sections.
4. Adjust the blood pressure monitoring system so that the output is 0 mmHg.
5. In the Blood Pressure section:
a. In the upper drop-down list select Manual Calibration.
b. Click Calibrate.
The blood pressure trace (red) should move to coincide with the 0 mark
on the blood pressure scale.
6. Adjust the blood pressure monitoring system to output a known level, and
note the numeric value of this level.
7. In the Blood Pressure section:
a. Set the BP Gain value to either 1X or 4X. The default value is 4X, which is
the typical setting for most devices.
b. Type the numeric value of the output level into the box.
c. Click Calibrate.
8. The system:
Calibrates your pressure scale.
Retains the calibration settings between imaging sessions. You only need
to repeat the calibration procedure if you connect a different blood
pressure monitor or if you think there might be a problem with the
calibration accuracy.
Respiration Gating section
Respiration gating is a tool you can use to effectively suppress the artifacts
coming from respiration and cardiac movement.
When you are acquiring image data along with physiological data, the physical
movement of the subject's chest cavity may move the region of interest you want
to study. This can cause artificial variations in measurements you add to saved
images.
Respiration gating suppresses this effect.
How respiration gating works
To suppress the effect of respiration on your image data, you use the Respiration
Gating tools to select the period of time between breaths – when the body is least
affected by the breathing motion. This brief period of time is called the respiration
gate. The system records image data only during the respiration gate period.
As shown in the following illustration, you work in the physiological trace
window to create the respiration gate along the yellow respiration data trace line.
The beginning of the respiration gate is called the delay point and the length of the
Vedere la pagina 114
1 2 ... 110 111 112 113 114 115 116 117 118 119 120 ... 452 453

Commenti su questo manuale

Nessun commento