Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 203
Chapter 34: Analyzing B-Mode images
f To place a linear distance measurement:
1. Click the linear distance measurement button .
2. Click on your image to place the initial caliper.
3. Trackball to the location where you want to end your measurement and then
click to place the end caliper. If you selected the Show Values and Labels
option in the Measurements tab of the Preferences window, the system
displays the measurement value and editable label for the measurement.
4. If you need to move the entire measurement, click on a measurement line and
then drag and drop it.
Related information
Complete procedure for adding a measurement (page 166)
Traced distance measurement
Traced distance is measured in mm.
f To place a traced distance measurement:
1. Click the traced distance measurement button .
2. Click on your image to place the initial caliper.
3. Trackball along the contour of your target tissue and then right-click to place
the final caliper of your trace. If you selected the Show Values and Labels
option in the Measurements tab of the Preferences window, the system
displays the measurement value and editable label for the measurement.
4. If you need to move the entire measurement, click on a measurement line and
then drag and drop it.
Related information
Complete procedure for adding a measurement (page 166)
2D Area measurement
2D Area is measured in mm
2
.
f To place a 2D area measurement:
1. Click the 2D area measurement button .
2. Click on your image to place the initial caliper.