332 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 45: Acquiring Power Doppler Mode images
Display
Parameter Description
Dynamic Range
The contrast of your image, measured in dB. Adjust with the
Dynamic Range
control.
Persistence
The state of the Persistence feature: Off, Low, Med, High, Max. Adjust with the
Persist
control.
Line Density
The line density level. One of four settings: Quarter, Third, Half, Full. Adjust with the
Line Density
control.
Display Map
The selected predefined display map from the predefined set of maps. Adjust with the
Display Map
control.
Wall Filter
The level of low velocity signals, measured in Hz, filtered out of the spectral display.
Adjust with the
Wall Filter
control.
Priority
The threshold level on the B-Mode gray scale, displayed as a percentage, above which
the system does not apply color data. Adjust with the
Priority
control.
Vedere la pagina 331