238 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 36: Analyzing M-Mode images
Refine the trace (page 238)
f To trace the outer LV walls as well as the inner LV walls:
Use the same peak and valley caliper points tracing method, but also trace the
outer LV walls using the following procedure:
1. On the upper wall, trace the outside wall along the number of cycles you want
to measure and then right-click to complete the trace. The outside wall is far
less dynamic than the inner wall.
2. On the upper wall, trace the inside peaks and valleys and right-click to
complete the trace.
3. On the lower wall, trace the inside peaks and valleys and right-click to
complete the trace.
4. On the lower wall, trace the outside wall and then right-click just once to
complete the trace and display the measurements.
Modifying the trace
You can modify the caliper points on your trace after you complete the trace.
f To move a caliper point:
1. Position the cursor over the point until the point becomes a pink cross.
2. Drag the point to a new position, then click again to set the new position.
f To add a point:
1. Position your cursor where you want to place a new point.
2. Click twice.
f To delete an individual point:
1. Position the cursor over the point until the point becomes a pink cross.
2. Right-click and select Delete Point. If more than one point is located within
the five-pixel radius, the system deletes the point that is closest point to the
cursor.
Refining the trace
After you complete a trace, you can use the system's Refine feature to
automatically adds points along a tissue layer to contour the trace more precisely
to the wall.
f To refine one trace line in your LV trace:
1. Right-click on the line and select Refine Current.
Vedere la pagina 237