Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 341
Chapter 47: Acquiring Contrast Mode images
The transducer transmits a single ultrasound pulse burst at maximum setting
for the period defined in the Contrast Mode preferences.
4. Press
Cine Store
.
Next steps
Displaying contrast agents as an overlay (page 348)
Adjusting the contrast overlay display (page 350)
Related information:
Typical B-Mode image acquisition session (page 190)
Typical Contrast 3D-Mode image acquisition
Contrast 3D-Mode adds Contrast Mode scan data during a 3D-Mode scan so you
can reconstruct a volume that integrates the Contrast Mode data with the
surrounding B-Mode 3D volume.
f To acquire a Contrast 3D-Mode image:
1. Set up for a 3D-Mode image acquisition session (page 282).
2. Press
Contrast
.
3. Complete the 3D motor stage initialization process and 3D acquisition setup
process as detailed in Typical 3D-Mode image acquisition session (page 275) and
click Scan.
The system acquires image slices across the motor stage track and combines
them into a cine loop. Unlike a typical cine loop which contains slices along
the same image plane over time, this cine loop contains a series of individual
slices at different locations as the motor stage moves along its track.
4. Inject the microbubbles according to the specified protocol and then press
3D
.
5. Press
Cine Store
to save the Contrast 3D-Mode image data.
6. Press
3D
.
The system acquires image slices at exactly the same step positions.
7. Click Destroy 3D.
The system stops acquiring data and runs the destruction level ultrasound
burst at each step along the the motor stage track and then returns the motor
stage to the initial position.
8. Press
3D
to acquire post-destruction image data.