254 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 37: Acquiring PW Doppler Mode and PW Tissue Doppler Mode images
Simul
This toggle control sets the system to acquire live data simultaneously in both the B-
Mode scout window as well as the PW Doppler image window.
In the dual window view, use this feature when you want to adjust your sample volume
in the B-Mode scout window while you view the waveform data in the PW Doppler
Mode window.
To use this toggle control:
1. Press to activate the simultaneous state.
A black vertical strip scans across the spectrum from left to right.
2. To eliminate this striping, press the toggle again to freeze the scout window and
return to PW Doppler image data only.
Active during: M-Mode, PW Doppler Mode and PW Tissue Doppler Mode image
acquisition sessions.
Sweep Speed
Adjusts the cine loop playback speed parameter so that you can stretch out or compress
the cine loop data in the review window. Push up to increase the speed and compress the
cine loop image. Pull down to decrease the speed and expand the cine loop image.
When you are reviewing the cine loop you can also use the
Cine Loop Review
control to
adjust the sweep speed.
In PW Doppler Mode and PW Tissue Doppler Mode: Set the sweep speed parameter in
a range from 0.25 seconds at 4000 Hz to 5.1 seconds at 200 Hz. In some cases, if your
imaging window is large and the
Velocity
is set high, the minimum speed may be greater.
The system displays the updated values in the status bar in the lower left area of the
screen.
Wall Filter
Filters out signals that correspond to low velocity axial motion. Typically these include
vessel wall movement, cardiac wall movement and tissue movement caused by
respiration. Push up to filter out more. Pull down to filter out less.
In PW Doppler Mode: Use this control to filter out the display of low velocity signal
artifacting that appears as a horizontal black band along either side of the white baseline.
Push up to reduce the lower velocity signals and bring the waveform of the spectral data
closer to the baseline. Pull down to display more low velocity signals.