102 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 16: Setting up your Vevo 2100 Imaging System
Storing the transducer
You can store the transducer in the transducer and gel holder attached to the side
of the Vevo 2100 Imaging System, nose upward and with the cable directed
toward the front of the cart.
Use the spring-loaded cable holder to ensure that the cable does not get twisted.
When you move the transducer from one facility to another, always use the
dedicated case that is provided with the cart.
f Follow these guidelines when you store the transducer in its case:
Make sure that the transducer is clean and dry before you store it in the case.
Place the transducer in the case carefully so the cable doesn't kink.
Don't store the transducer in areas of extreme temperatures or in direct
sunlight.
Store the transducer separately from other instruments so it won't get
damaged accidentally.
Related information
Front view of the Vevo 2100 Imaging System (page 17)
Working with the 3D motor stage (optional)
VisualSonics provides a 3D motor stage for customers who need to perform 3D
volumetric measurements. The 3D motor stage connects to the Vevo Imaging
Station.
IMPORTANT: During 3D data acquisition, ensure that the animal under the transducer
is flat in relation to the 3D scan direction to prevent unintended contact with the animal
when the transducer moves.
Connecting the 3D motor stage to the Vevo Imaging Station
The 3D motor stage features a Quick Release post on the top to connect to the
Vevo Imaging Station, and a Quick Release mount on the bottom to affix the
transducer clamp.