Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 339
Chapter 47: Acquiring Contrast Mode images
3. Acquire 100 to 200 frames of data and then save and label the cine loop as
Baseline.
4. Press
Pre Trigger
and inject the contrast agent. If you selected Auto SAVE on
Scan Completion for Contrast Mode in the General tab of the Preferences
window the system saves the cine loop when the acquisition ends.
You have created a cine loop of the bolus injection.
f To automatically create a contrast agent destruction cine loop:
1. Press
Contrast
and begin acquiring image data.
2. Position the transducer and locate your region of interest.
3. Inject the contrast agent according to your protocol and then press
Image Sequence
.
The system completes an automated sequence of actions based on the
configuration you define in the Contrast Mode (page 74) section in the
General tab of the Preferences window:
a. The system acquires data for a set portion of the default cine loop length
as you inject the contrast agent.
b. The transducer transmits a single ultrasound pulse at maximum setting
for a short specified period. This destroys the contrast agent in the region
of interest.
c. The system acquires data for the remainder of the cine loop.
4. Press
Cine Store
.