
Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 117
Chapter 18: Setting up to acquire physiological data
The left panel displays the physiological display setting sections.
3. In the Physiological Range section, adjust the Respiration slider so that the
trace line is a) short enough that the peaks and valleys do not extend above or
below the window and b) tall enough that you can clearly define those peaks
and valleys.
4. In the Respiration Gating section:
a. Select the Respiration Gating check box to activate the slider controls.
b. Adjust the Delay slider to set the start of the gate period, after the
waveform has returned to the baseline.
c. Adjust the Window slider to set the duration of the data acquisition before
the next breath occurs.
5. Press
Pre Trigger
to create your cine loop.
Because
Pre Trigger
records data for a set period after you press the key, the
system acquires only a portion of data during each cardiac cycle, so it takes longer
to acquire the cine loop.
Related information
Acquiring image data (page 120)
Physiological data sources (page 109)
ECG Trigger section (page 117)
ECG Trigger section
ECG triggering is a feature you can use to effectively acquire imaging frames at a
specific time during the heart cycle.
ECG triggering suppresses this effect.
Use ECG triggering when you intend to add measurements at a specific time.
Commenti su questo manuale